1.
Sci Rep ; 14(1): 9557, 2024 04 25.
Artigo em Inglês | MEDLINE | ID: mdl-38664481

RESUMO

Breakthrough multi-response miniature dosimetry/spectrometry of electroneutrons (EN) was made on surface and in-depths of whole-body polyethylene phantom under 10 cm × 10 cm electron beam of 20 MV Varian Clinac 2100C electron medical accelerator commonly applied for prostate treatment. While dosimetry/spectrometry of photoneutrons (PN) has been well characterized for decades, those of ENs lagged behind due to very low EN reaction cross section and lack of sensitive neutron dosimeters/spectrometers meeting neutron dosimetry requirements. Recently, Sohrabi "miniature neutron dosimeter/spectrometer" and "Stripe polycarbonate dosimeter" have broken this barrier and determined seven EN ambient dose equivalent (ENDE) (µSv.Gy-1) responses from electron beam and from albedo ENs including beam thermal (21 ± 2.63), albedo thermal (43 ± 3.70), total thermal (64 ± 6.33), total epithermal (32 ± 3.90), total fast (112.00), total thermal + epithermal (l96 ± 10), and total thermal + epithermal + fast (208 ± 10.23) ENs. Having seven ENDE responses of this study and seven PNDE responses of previous study with the same accelerator obtained at identical conditions by the same principle author provided the opportunity to compare the two sets of responses. The PNDE (µSv.Gy-1) responses have comparatively higher values and 22.60 times at isocenter which provide for the first time breakthrough ENDE responses not yet reported in any studies before worldwide.


Assuntos
Aceleradores de Partículas , Radiometria , Aceleradores de Partículas/instrumentação , Radiometria/instrumentação , Radiometria/métodos , Nêutrons , Humanos , Elétrons , Imagens de Fantasmas
2.
PLoS One ; 19(4): e0301435, 2024.
Artigo em Inglês | MEDLINE | ID: mdl-38635642

RESUMO

In radiotherapy, when photon energy exceeding 8 MV is utilized, photoneutrons can activate the components within the gantry of the linear accelerator (linac). At the end of the linac's lifecycle, radiation workers are tasked with its dismantling and disposal, potentially exposing them to unintentional radiation. This study aims to identify and measure the radioisotopes generated by this activation through spectroscopy, and to evaluate the effective dose rate. We selected nine medical linacs, considering various factors such as manufacturer (Siemens, Varian, and Elekta), model, energy, period of operation, and workload. We identified the radionuclides in the linac head by employing an in situ high-purity germanium (HPGe) detector. Spectroscopy and dose-rate measurements were conducted post-shutdown. We also measured the dose rates at the beam-exit window following irradiation with 10 MV and 15 MV photon beams. As a result of the spectroscopy, we identified approximately 20 nuclides including those with half-lives of 100 days or longer, such as 54Mn, 60Co, 65Zn, 122Sb, and 198Au. The dose rate measurements after 10 MV irradiation decreased to the background level in 10 min. By contrast, on 15 MV irradiation, the dose rate was 628 nSv/h after 10 min and decreased to 268 nSv/h after 1.5 hours. It was confirmed that the difference in the level of radiation and the type of nuclide depends on the period of use, energy, and workload. However, the type of nuclide does not differ significantly between the linacs. It is necessary to propose appropriate guidelines for the safety of workers, and disposal/move-install should be planned while taking into consideration the equipment's energy usage rate.


Assuntos
Manganês , Radioisótopos , Humanos , Dosagem Radioterapêutica , Aceleradores de Partículas , Fótons , Análise Espectral
4.
Zhongguo Yi Liao Qi Xie Za Zhi ; 48(2): 156-159, 2024 Mar 30.
Artigo em Chinês | MEDLINE | ID: mdl-38605614

RESUMO

Objective: The distribution of the photon energy spectrum in isocenter plane of the medical linear accelerator and the influence of secondary collimator on the photon energy spectrum are studied. Methods Use the BEAMnrc program to simulate the transmission of the 6 MeV electrons and photons in 5 cm×5 cm,10 cm×10 cm,15 cm×15 cm and 20 cm×20 cm fields in treatment head of the medical linear accelerator, where a phase space file was set up at the isocenter plane to record the particle information passing through this plane. The BEAMdp program is used to analyze the phase space file, in order to obtain the distribution of the photon energy spectrum in isocenter plane and the influence of secondary collimator on the photon energy spectrum. Results: By analyzing the photon energy spectrum of a medical linear accelerator with a nominal energy of 6 MV, it is found that the secondary collimator has little effect on the photon energy spectrum; different fields have different photon energy spectrum distributions; the photon energy spectrum in different central regions of the same field have the same normalized distribution. Conclusion: In the dose calculation of radiation therapy, the influence of photon energy spectrum should be carefully considered.


Assuntos
Fótons , Planejamento da Radioterapia Assistida por Computador , Método de Monte Carlo , Fótons/uso terapêutico , Aceleradores de Partículas , Imagens de Fantasmas , Dosagem Radioterapêutica
5.
Zhongguo Yi Liao Qi Xie Za Zhi ; 48(2): 212-216, 2024 Mar 30.
Artigo em Chinês | MEDLINE | ID: mdl-38605624

RESUMO

The medical electron linear accelerator(LINAC) has the characteristics of complex system structure, many core components and high precision control requirements, which puts forward higher requirements for product quality control and regulation. This study puts forward the main points of field inspection through the analysis of the technical characteristics and production risk of LINAC, combined with the requirements of the good manufacturing practice of medical devices. It has certain reference significance for quality management personnel and field inspectors.


Assuntos
Elétrons , Aceleradores de Partículas
6.
Zhongguo Yi Liao Qi Xie Za Zhi ; 48(2): 184-191, 2024 Mar 30.
Artigo em Chinês | MEDLINE | ID: mdl-38605619

RESUMO

More than 70% of tumor patients require radiotherapy. Medical electron linear accelerators are important high-end radiotherapy equipment for tumor radiotherapy. With the application of artificial intelligence technology in medical electron linear accelerator, radiotherapy has evolved from ordinary radiotherapy to today's intelligent radiotherapy. This study introduces the development history, working principles and system composition of medical electron linear accelerators. It outlines the key technologies for improving the performance of medical linear electron accelerators, including beam control, multi-leaf collimator, guiding technology and dose evaluation. It also looks forward to the development trend of major radiotherapy technologies, such as biological guided radiotherapy, FLASH radiotherapy and intelligent radiotherapy, which provides references for the development of medical electron linear accelerators.


Assuntos
Elétrons , Neoplasias , Humanos , Inteligência Artificial , Aceleradores de Partículas , Dosagem Radioterapêutica
7.
BMC Cancer ; 24(1): 324, 2024 Mar 08.
Artigo em Inglês | MEDLINE | ID: mdl-38459443

RESUMO

BACKGROUND AND PURPOSE: Radiotherapy (RT) is an essential treatment modality against cancer and becoming even more in demand due to the anticipated increase in cancer incidence. Due to the rapid development of RT technologies amid financial challenges, we aimed to assess the available RT facilities and the issues with achieving health equity based on current equipment compared to the previous reports from Iran. MATERIALS AND METHODS: A survey arranged by the Iran Cancer Institute's Radiation Oncology Research Center (RORC) was sent to all of the country's radiotherapy centers in 2022. Four components were retrieved: the reimbursement type, equipment, human resources, and patient load. To calculate the radiotherapy utilization rate (RUR), the Lancet Commission was used. The findings were compared with the previous national data. RESULTS: Seventy-six active radiotherapy centers with 123 Linear accelerators (LINACs) were identified. The centers have been directed in three ways. 10 (20 LINACs), 36 (50 LINACs), and 30 centers (53 LINACs) were charity-, private-, and public-based, respectively. Four provinces had no centers. There was no active intraoperative radiotherapy machine despite its availability in 4 centers. One orthovoltage X-ray machine was active and 14 brachytherapy devices were treating patients. There were 344, 252, and 419 active radiation oncologists, medical physicists, and radiation therapy technologists, respectively. The ratio of LINAC and radiation oncologists to one million populations was 1.68 and 4.10, respectively. Since 2017, 35±5 radiation oncology residents have been trained each year. CONCLUSION: There has been a notable growth in RT facilities since the previous reports and Iran's situation is currently acceptable among LMICs. However, there is an urgent need to improve the distribution of the RT infrastructure and provide more facilities that can deliver advanced techniques.


Assuntos
Neoplasias , Radioterapia (Especialidade) , Humanos , Irã (Geográfico)/epidemiologia , Neoplasias/epidemiologia , Neoplasias/radioterapia , Aceleradores de Partículas , Inquéritos e Questionários , Radioterapia/métodos
8.
Cancer Radiother ; 28(2): 164-173, 2024 Apr.
Artigo em Francês | MEDLINE | ID: mdl-38490925

RESUMO

PURPOSE: Radiosurgery for the treatment of trigeminal neuralgia delivers a very high dose in a single fraction, over a few millimeters, at a single isocenter placed along the nerve. We present here the different steps that have been performed to validate small beams by conical collimators, and report the clinical results of the first patients treated on Novalis Tx®, frameless. MATERIAL AND METHODS: First, the geometric accuracy of 4 and 6mm conical collimators was evaluated using Winston-Lutz tests; then dosimetric data acquisition was performed using high spatial resolution detectors (PTW 60019 microdiamond and a PTW 60017 E-diode). The corrective factors of the TRS 483 report were applied to calculate the collimator aperture factors. These dosimetric data were then compared with the data implemented in the iPlan® treatment planning system. Then end-to-end tests were performed to control the entire treatment process using an anthropomorphic phantom "STEEV". Between 2020 and 2022, 18 patients were treated for refractory trigeminal neuralgia on Novalis Tx®, frameless, with Exactrac® repositioning. A total of 17 patients were evaluated (one was lost to follow-up) using the BNI score for pain assessment and MRI with a median follow-up of 12 months. RESULTS: The quality criteria of geometric and dosimetric accuracy were met for the 6-mm cone but not for the 4-mm cone. All patients were treated with a 6-mm cone with a dose of 90Gy prescribed at the isocenter at the root entry zone. Initial pain control was obtained in 70.5% of our patients, and 53% maintained pain control with a median follow-up of 12 months. All recurrences occurred within 3 to 6 months after radiosurgery. No brainstem toxicity was observed. Six patients had non-disabling facial hypoesthesia, half of whom already had pretreatment hypoesthesia. CONCLUSION: The treatment of trigeminal neuralgia on a dedicated linear accelerator is a highly technical treatment whose accuracy and safety are paramount. The physical measurements allowed the commissioning of the technique with a 6mm cone. Our first clinical results are in accordance with the literature.


Assuntos
Radiocirurgia , Neuralgia do Trigêmeo , Humanos , Radiocirurgia/métodos , Neuralgia do Trigêmeo/radioterapia , Hipestesia/cirurgia , Aceleradores de Partículas , Dor/cirurgia , Resultado do Tratamento
9.
J Cancer Res Ther ; 20(1): 224-231, 2024 Jan 01.
Artigo em Inglês | MEDLINE | ID: mdl-38554325

RESUMO

PURPOSE: The purpose of this study is to evaluate the dosimetric and treatment delivery characteristics of volumetric modulated arc therapy technique (VMAT)-based craniospinal axis irradiation (CSI) between ring gantry Halcyon (HAL) and C-arm based Novalis Tx (NTx) linear accelerator. Set-up margin and treatment delivery time for both machines were also taken into account. MATERIALS AND METHODS: Fifteen patients, 4 females and 11 males treated between March 2019 and February 2022 within the age group 4-56 years simulated in the supine position and were planned for multiple isocentre VMAT technique in ring gantry Halcyon and C-Arm Novalis linear accelerator for 6FFF and 6 MV flatten beam energy. The number of isocenters was the same in both the machines, usually three for adult adolescent age group patients and two for pediatric patients. Total on-couch time and the patient positional shift were captured for each isocenter during each session of treatment. Margins were calculated using Herk's formula of margin = 2.5Σ +0.7σ. Dosimetry, on-couch time, and set-up margin were compared between two competing arms. RESULTS: Ninety-five percent of PTV coverage (P = 0.333), volume receiving 107% (P = 0.676), total MU (P = 0.818) in both the arms were comparable and statically insignificant. Low-dose spillage such as D20% (P = 0.212) and D50% (P = 0.008) was lesser in HAL comparable to NTx. CI and HI were statically insignificant. Out of 26 organs at risk (OAR), only 3 organs showed a statically significant dose difference. The mean and maximum setup margin in any linear direction was 0.45 and 0.53 cm for HAL and 0.37 and 0.56 cm for NTx and, variation was statistically insignificant (0.23 < P < 0.47). On-couch time was 4.0 ± 5.5 min lesser for HAL and the difference in on-couch time between the two arms was statistically different. CONCLUSION: Even though the majority of the delivery parameters such as gantry speed, dose rate, beam characteristic (flatten or unflatten), MLC width, and speed between the ring gantry HAL and C-arm NTx linear accelerators were distinctly different, they offered no or minimal difference in the dose distribution and in the setup margin. HAL gives a faster treatment time delivery, which could be crucial for some selective cases such as patients receiving treatment under general anesthesia.


Assuntos
Radiação Cranioespinal , Radioterapia de Intensidade Modulada , Masculino , Adulto , Feminino , Humanos , Criança , Adolescente , Pré-Escolar , Adulto Jovem , Pessoa de Meia-Idade , Radioterapia de Intensidade Modulada/métodos , Planejamento da Radioterapia Assistida por Computador/métodos , Dosagem Radioterapêutica , Aceleradores de Partículas
10.
J Cancer Res Ther ; 20(1): 389-395, 2024 Jan 01.
Artigo em Inglês | MEDLINE | ID: mdl-38554351

RESUMO

INTRODUCTION: Total skin electron beam therapy, commonly known as TSET, is a good choice of treatment for patients suffering from mycosis fungoides. The aim of this study was to introduce a new approach to the beam profile measurement using diodes and to calculate the monitor units required for the TSET treatment by the use of a simple setup of output measurement. Dosimetric measurements required for the treatment were taken to establish the Stanford technique in the department, and the measured data was compared with the published data. MATERIALS AND METHODS: High-energy Linear Accelerator Clinac-DHX, Varian medical system, Palo Alto, CA, was commissioned for TSET. The output of the machine was measured by the use of a Parallel-Plate Chamber (PPC40) as per the TRS 398 recommendation. Diode dosimeters (EDD2 and EDD5) were used for beam profile measurements due to easy setup and to reduce the measurement time. RESULTS: Homogeneous dose distribution within a field size of 80 cm x160 cm was observed with the variation of -5.0% on the horizontal axis and -5.4% on the vertical axis. The calculated monitor unit to deliver 200 cGy per fraction per field at the source to surface (SSD) of 416 cm was 489 MU. CONCLUSION: The technique described for the output measurements is simple and accurate. Results of the absorbed dose and MU measured were within good agreement compared to the published literature.


Assuntos
Aceleradores de Partículas , Radiometria , Humanos , Dosagem Radioterapêutica , Radiometria/métodos
11.
Sci Rep ; 14(1): 7134, 2024 03 26.
Artigo em Inglês | MEDLINE | ID: mdl-38532018

RESUMO

We aimed to investigate the deliverability of dynamic conformal arc therapy (DCAT) by gantry wobble owing to the intrinsic inter-segment break of the Elekta linear accelerator (LINAC) and its adverse influence on the dose to the patient. The deliverability of DCAT was evaluated according to the plan parameters, which affect the gantry rotation speed and resultant positional inaccuracies; the deliverability according to the number of control points and dose rates was investigated by using treatment machine log files and dosimetry devices, respectively. A non-negligible degradation in DCAT deliverability due to gantry wobble was observed in both the treatment machine log files and dosimetry devices. The resulting dose-delivery error occurred below a certain number of control points or above a certain dose rate. Dose simulations in the patient domain showed a similar impact on deteriorated deliverability. For targets located primarily in the isocenter, the dose differences were negligible, whereas for organs at risk located mainly off-isocenter, the dose differences were significant up to - 8.77%. To ensure safe and accurate radiotherapy, optimal plan parameters should be selected, and gantry angle-specific validations should be conducted before treatment.


Assuntos
Radioterapia Conformacional , Radioterapia de Intensidade Modulada , Humanos , Dosagem Radioterapêutica , Radioterapia Conformacional/métodos , Planejamento da Radioterapia Assistida por Computador/métodos , Aceleradores de Partículas , Radiometria/métodos , Radioterapia de Intensidade Modulada/métodos
12.
J Appl Clin Med Phys ; 25(4): e14323, 2024 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-38426612

RESUMO

The Elekta Unity magnetic resonance (MR) linac is limited to longitudinal couch motion and a sagittal-only laser, which restricts the ability to perform patient-specific quality assurance (PSQA) intensity-modulated radiotherapy (IMRT) measurements for very lateral targets. This work introduces a simple method to perform PSQA using the Sun Nuclear ArcCheck-MR phantom at left and right lateral positions without additional equipment or in-house construction. The proposed setup places the center of the phantom 1.3 cm vertical and 12.9 cm lateral to isocenter in either the left or right direction. Computed tomography (CT) scans are used to simulate the setup and create a QA plan template in the Monaco treatment planning system (TPS). The workflow is demonstrated for four patients, with an average axial distance from the center of the bore to the planning target volume (PTV) of 12.4 cm. Gamma pass rates were above 94% for all plans using global 3%/2 mm gamma criterion with a 10% threshold. Setup uncertainties are slightly larger for the proposed lateral setup compared to the centered setup on the Elekta platform (∼1 mm compared to ∼0.5 mm), but acceptable pass rates are achievable without optimizing shifts in the gamma analysis software. In general, adding the left and right lateral positions increases the axial area in the bore encompassed by the cylindrical measurement array by 147%, substantially increasing the flexibility of measurements for offset targets. Based on this work, we propose using the lateral QA setup if the closest distance to the PTV edge from isocenter is larger than the array radius (10.5 cm) or the percent of the PTV encompassed by the diode array would be increased with the lateral setup compared to the centered setup.


Assuntos
Planejamento da Radioterapia Assistida por Computador , Radioterapia de Intensidade Modulada , Humanos , Planejamento da Radioterapia Assistida por Computador/métodos , Imageamento por Ressonância Magnética , Aceleradores de Partículas , Radioterapia de Intensidade Modulada/métodos , Espectroscopia de Ressonância Magnética , Dosagem Radioterapêutica
13.
Radiat Prot Dosimetry ; 200(6): 538-543, 2024 Apr 20.
Artigo em Inglês | MEDLINE | ID: mdl-38441907

RESUMO

The objective of this study is dosimetric comparison between the O-ring Halcyon and C-arm Clinac iX linac for volumetric modulated arc therapy (VMAT) plans for head & neck (H&N) cancer and carcinoma cervix patients. Total 60 patients of H&N cancer and carcinoma cervix were enrolled prospectively from March 2021 to March 2023. VMAT plans with 6 MV photons for Halcyon and Clinac iX were generated and compared for each patient by dose volume histogram for planning target volume coverage and organ at risk (OAR) sparing. There were no differences in between both the linacs for PTV D2% and D98%, homogeneity index, conformity index, Dmax (maximum dose) and Dmean (mean dose) of OAR. Halcyon had significantly shorter treatment time compared to Clinac iX. Halcyon delivered higher integral dose and monitor units. O-ring Halcyon produces VMAT plans comparable to other C-arm linacs for H&N and carcinoma cervix patients.


Assuntos
Neoplasias de Cabeça e Pescoço , Órgãos em Risco , Aceleradores de Partículas , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador , Radioterapia de Intensidade Modulada , Neoplasias do Colo do Útero , Humanos , Radioterapia de Intensidade Modulada/métodos , Feminino , Neoplasias do Colo do Útero/radioterapia , Planejamento da Radioterapia Assistida por Computador/métodos , Aceleradores de Partículas/instrumentação , Neoplasias de Cabeça e Pescoço/radioterapia , Órgãos em Risco/efeitos da radiação , Radiometria/métodos , Estudos Prospectivos , Pessoa de Meia-Idade , Adulto
14.
Radiat Oncol ; 19(1): 40, 2024 Mar 20.
Artigo em Inglês | MEDLINE | ID: mdl-38509543

RESUMO

PURPOSE: To assess a large panel of MR compatible detectors on the full range of measurements required for a 0.35 T MR-linac commissioning by using a specific statistical method represented as a continuum of comparison with the Monte Carlo (MC) TPS calculations. This study also describes the commissioning tests and the secondary MC dose calculation validation. MATERIAL AND METHODS: Plans were created on the Viewray TPS to generate MC reference data. Absolute dose points, PDD, profiles and output factors were extracted and compared to measurements performed with ten different detectors: PTW 31010, 31021, 31022, Markus 34045 and Exradin A28 MR ionization chambers, SN Edge shielded diode, PTW 60019 microdiamond, PTW 60023 unshielded diode, EBT3 radiochromic films and LiF µcubes. Three commissioning steps consisted in comparison between calculated and measured dose: the beam model validation, the output calibration verification in four different phantoms and the commissioning tests recommended by the IAEA-TECDOC-1583. MAIN RESULTS: The symmetry for the high resolution detectors was higher than the TPS data of about 1%. The angular responses of the PTW 60023 and the SN Edge were - 6.6 and - 11.9% compared to the PTW 31010 at 60°. The X/Y-left and the Y-right penumbras measured by the high resolution detectors were in good agreement with the TPS values except for the PTW 60023 for large field sizes. For the 0.84 × 0.83 cm2 field size, the mean deviation to the TPS of the uncorrected OF was - 1.7 ± 1.6% against - 4.0 ± 0.6% for the corrected OF whereas we found - 4.8 ± 0.8% for passive dosimeters. The mean absolute dose deviations to the TPS in different phantoms were 0 ± 0.4%, - 1.2 ± 0.6% and 0.5 ± 1.1% for the PTW 31010, PTW 31021 and Exradin A28 MR respectively. CONCLUSIONS: The magnetic field effects on the measurements are considerably reduced at low magnetic field. The PTW 31010 ionization chamber can be used with confidence in different phantoms for commissioning and QA tests requiring absolute dose verifications. For relative measurements, the PTW 60019 presented the best agreement for the full range of field size. For the profile assessment, shielded diodes had a behaviour similar to the PTW 60019 and 60023 while the ionization chambers were the most suitable detectors for the symmetry. The output correction factors published by the IAEA TRS 483 seem to be applicable at low magnetic field pending the publication of new MR specific values.


Assuntos
Aceleradores de Partículas , Radiometria , Humanos , Radiometria/métodos , Método de Monte Carlo , Imagens de Fantasmas , Calibragem
15.
Technol Cancer Res Treat ; 23: 15330338241239144, 2024.
Artigo em Inglês | MEDLINE | ID: mdl-38515394

RESUMO

Background: This review investigates peripheral dose levels in electron beam treatments, comparing different manufacturers including Varian, Elekta, and Siemens. Accurate measurement of peripheral dose is vital for patient safety and precise radiation delivery in radiation therapy. Methods: This review followed PRISMA standards, conducting a comprehensive literature search from 1978 to July 2023. Emphasis was on identifying studies analyzing peripheral doses related to various electron beam energies, beam angle, field sizes, cutouts, and applicator combinations. Three major databases including PubMed, Web of Science, and Scopus were searched. Results: A total of 7 articles were included in this review. Strategies such as bolus materials, personalized cutouts, and optimal treatment procedures have all been developed to reduce peripheral radiation exposure and enhance patient safety. Ongoing research in this field is focused on further minimizing the risks associated with out-of-field radiation by improving dose delivery systems. Conclusion: The literature emphasizes importance of precision in electron beam radiation therapy, highlighting the critical need for managing peripheral doses and optimizing hardware to ensure patient safety. It advocates for the use of advanced tools and protocols to maintain a balance between effective treatment while protecting healthy tissues. Continuous research, careful treatment planning, and effective management of peripheral doses are essential.


Assuntos
Elétrons , Planejamento da Radioterapia Assistida por Computador , Humanos , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Imagens de Fantasmas , Aceleradores de Partículas
16.
Radiat Res ; 201(3): 252-260, 2024 Mar 01.
Artigo em Inglês | MEDLINE | ID: mdl-38308528

RESUMO

This study aims to investigate the feasibility of enhancing the charge collection efficiency (CCE) of a transmission chamber by reconfiguring its design and operation. The goal was to extend the range of dose-per-pulse (DPP) values with no or minimal recombination effects up to the ultra-high dose rate (UHDR) regime. The response of two transmission chambers, with electrode distance of 1 mm and 0.6 mm, respectively, was investigated as a function of applied voltage. The chambers were mounted one-by-one in the electron applicator of a 10 MeV FLASH-modified clinical linear accelerator. The chamber signals were measured as a function of nominal DPP, which was determined at the depth of dose maximum using EBT-XD film in solid water and ranged from 0.6 mGy per pulse to 0.9 Gy per pulse, for both the standard voltage of 320 V and the highest possible safe voltage of 1,200 V. The CCE was calculated and fitted with an empirical logistic function that incorporated the electrode distance and the chamber voltage. The CCE decreased with increased DPP. The CCE at the highest achievable DPP was 24% (36%) at 320 V and 51% (82%) at 1,200 V, for chambers with 1 mm (0.6 mm) electrode distance. For the combination of 1,200 V- and 0.6-mm electrode distance, the CCE was ∼100% for average dose rate up to 70 Gy/s at the depth of dose maximum in the phantom at a source-to-surface distance of 100 cm. Our findings indicate that minor modifications to a plane-parallel transmission chamber can substantially enhance the CCE and extending the chamber's operating range to the UHDR regime. This supports the potential of using transmission chamber-based monitoring solutions for UHDR beams, which could facilitate the delivery of UHDR treatments using an approach similar to conventional clinical delivery.


Assuntos
Aceleradores de Partículas , Radiometria , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador , Imagens de Fantasmas
17.
Phys Med Biol ; 69(7)2024 Mar 14.
Artigo em Inglês | MEDLINE | ID: mdl-38412538

RESUMO

Objective.The Elekta unity MR-linac delivers step-and-shoot intensity modulated radiotherapy plans using a multileaf collimator (MLC) based on the Agility MLC used on conventional Elekta linacs. Currently, details of the physical Unity MLC and the computational model within its treatment planning system (TPS)Monacoare lacking in published literature. Recently, a novel approach to characterize the physical properties of MLCs was introduced using dynamic synchronous and asynchronous sweeping gap (aSG) tests. Our objective was to develop a step-and-shoot version of the dynamic aSG test to characterize the Unity MLC and the computational MLC models in theMonacoandRayStationTPSs.Approach.Dynamic aSG were discretized into a step-and-shoot aSG by investigating the number of segments/sweep and the minimal number of monitor units (MU) per segment. The step-and-shoot aSG tests were compared to the dynamic aSG tests on a conventional linac at a source-to-detector distance of 143.5 cm, mimicking the Unity configuration. the step-and-shoot aSG tests were used to characterize the Unity MLC through measurements and dose calculations in both TPSs.Main results.The step-and-shoot aSGs tests with 100 segments and 5 MU/segment gave results very similar to the dynamic aSG experiments. The effective tongue-and-groove width of the Unity gradually increased up to 1.4 cm from the leaf tip end. The MLC models inRayStationandMonacoagreed with experimental data within 2.0% and 10%, respectively. The largest discrepancies inMonacowere found for aSG tests with >10 mm leaf interdigitation, which are non-typical for clinical plans.Significance.The step-and-shoot aSG tests accurately characterize the MLC in step-and-shoot delivery mode. The MLC model inRayStation2023B accurately describes the tongue-and-groove and leaf tip effects whereasMonacooverestimates the tongue-and-groove shadowing further away from the leaf tip end.


Assuntos
Planejamento da Radioterapia Assistida por Computador , Radioterapia de Intensidade Modulada , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia de Intensidade Modulada/métodos , Aceleradores de Partículas , Radiometria/métodos
18.
J Appl Clin Med Phys ; 25(3): e14291, 2024 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-38306504

RESUMO

PURPOSE: To present a modified calibration method to reduce signal drift due to table sagging in Respiratory Gating for Scanner (RGSC) systems with a wall-mounted camera. MATERIALS AND METHODS: Approximately 70 kg of solid water phantoms were evenly distributed on the CT couch, mimicking the patient's weight. New calibration measurements were performed at 9 points at the combination of three lateral positions, the CT isocenter and ±10 cm laterally from the isocenter, and three longitudinal locations, the CT isocenter and ±30 cm or ±40 cm from the isocenter. The new calibration was tested in two hospitals. RESULTS: Implementing the new weighed calibration method at the extended distance yielded improved results during the DIBH scan, reducing the drift to within 1 from 3 mm. The extended calibration positions exhibited similarly reduced drift in both hospitals, reinforcing the method's robustness and its potential applicability across all centers. CONCLUSION: This proposed solution aims to minimize the systematic error in radiation delivery for patients undergoing motion management with wall-mounted camera RGSC systems, especially in conjunction with a bariatric CT couchtop.


Assuntos
Aceleradores de Partículas , Humanos , Imagens de Fantasmas , Movimento (Física)
19.
Phys Med Biol ; 69(6)2024 Mar 13.
Artigo em Inglês | MEDLINE | ID: mdl-38408387

RESUMO

Objective. Real-time MRgRT uses 2D-cine imaging for target tracking and motion evaluation. Rotation of gantry inducedB0off-resonance, resulting in image artifacts and imaging isocenter-shift precluding MR-guided arc therapy. Standard MRI phantoms designed for higher resolution images face challenges when low-resolution cine imaging is needed to achieve high frame rates. This work aimed to examine the spatial accuracy including geometric distortion and isocenter shift in real-time during gantry rotation on a 0.35 T MR-Linac using the concentric Cine imaging quality assurance (QA) phantom and its associated image analysis software.Approach. The Cine imaging QA phantom consists of two concentric shells of low-T1mineral oil and a central alignment structure. The phantom was scanned on three different MRI systems; 0.55 T Siemens Free.Max, 1.5 T Philips Ingenia, and 0.35 T ViewRay MRIdian MR-Linac using 2D balanced steady-state free precession (bSSFP) imaging sequence. In addition, bSSFP cine MRI with the banding artifact correction was tested on 0.35 T ViewRay MR-Linac. Images from the MR-Linac were acquired with the Linac gantry stationary and rotating from gantry 300°â†’ 0° and vice versa. Three orthogonal image planes were scanned excluding the 1.5 T Philips Ingenia, where only the axial plane was scanned. The image analysis software calculated the distortion values as well as the isocenter position for each cine frame.Main results. The geometric distortion of cine imaging on MRIs and MR-Linac at gantry stationary are within 1 mm while the substantial geometric distortion of 2 and 2.2 mm were observed on 0.35 T MR-Linac while rotating the gantry clockwise (300°â†’ 0°) and counterclockwise 0°â†’ 300° respectively. The average imaging isocenter shift was 0.1 mm for both MRIs and the static gantry and imaging isocenter shift of ≤1.5 mm was observed during the gantry rotation. The imaging isocenter shift decreased by 1 ± 0.2 mm clockwise and counterclockwise withB0compensation.Significance. The concentric Cine imaging QA phantom and its associated software effectively demonstrate the image distortion on real-time cine imaging on regular MRIs and 0.35 T MR-Linac. The results of significant geometric distortion with a rotating gantry in the MR-Linac system require further investigation to alleviate the extent of the image distortion.


Assuntos
Processamento de Imagem Assistida por Computador , Aceleradores de Partículas , Processamento de Imagem Assistida por Computador/métodos , Software , Imagens de Fantasmas , Imageamento por Ressonância Magnética/métodos
20.
Med Phys ; 51(3): 2155-2163, 2024 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-38308857

RESUMO

BACKGROUND: MR-LINAC systems have been increasingly utilized for real-time imaging in adaptive treatments worldwide. Challenges in MR representation of air cavities and subsequent estimation of electron density maps impede planning efficiency and may lead to dose calculation uncertainties. PURPOSE: To demonstrate the generation of accurate electron density maps using the primary MV beam with a flat-panel imager. METHODS: The ViewRay MRIdian MR-LINAC system was modeled digitally for Monte Carlo simulations. Iron shimming, the magnetic field, and the proposed flat panel detector were included in the model. The effect of the magnetic field on the detector response was investigated. Acquisition of projections over 360 degrees was simulated for digital phantoms of the Catphan 505 phantom and a patient treated for Head and Neck cancer. Shim patterns on the projections were removed and detector noise linearity was assessed. Electron density maps were generated for the digital patient phantom using the flat-panel detector and compared with actual treatment planning CT generated electron density maps of the same patient. RESULTS: The effect of the magnetic field on the detector point-spread function (PSF) was found to be substantial for field strengths above 50 mT. Shims correction in the projection images using air normalization and in-painting effectively removed reconstruction artifacts without affecting noise linearity. The relative difference between reconstructed electron density maps from the proposed method and electron density maps generated from the treatment planning CT was 11% on average along all slices included in the iMREDe reconstruction. CONCLUSIONS: The proposed iMREDe technique demonstrated the feasibility of generating accurate electron densities for the ViewRay MRIdian MR-LINAC system with a flat-panel imager and the primary MV beam. This work is a step towards reducing the time and effort required for adaptive radiotherapy in the current ViewRay MR-LINAC systems.


Assuntos
Elétrons , Neoplasias de Cabeça e Pescoço , Humanos , Tomografia Computadorizada de Feixe Cônico/métodos , Imagens de Fantasmas , Método de Monte Carlo , Aceleradores de Partículas
SELEÇÃO DE REFERÊNCIAS